MODELiPulse
AUTO/MANUAL MODE Automated, manual
POWER NEEDED, VAC 100-230
WEIGHT, kg 57 (125.6), cart 29 (63.9)
DIMENSIONS 86 x 58 x 28 (33.8 x 22.8 x 11)
APPLICATIONS Adult
SIGNAL DATA ECG, pressure
DATA MANAGEMENT
PRINTER Via strip recorder
Events stored Alarms, power loss, time
Storage device Flash memory, Ethernet
DISPLAY
Type/size 12.1“ 1024 x 768 high-resolution LCD SVGA
Qrs indicator Yes
Pressure alert Yes
Lead select 9-lead
PARAMETERSECG, arterial pressure, balloon pressure, timing mode, trigger source, ECG source, ECG gain, AP source, helium level, battery status, AC power status, IAB augmentation level, heart rate, AP (systolic, diastolic, mean, augmented), augmentation alarm, R-wave deflation indicator, alarm mute, inflation/deflation markers
STRIP RECORDER Yes
AUTO CONDENSATION REMOVALFully automatic removal and disposal
TRIGGERING MECHANISMS ECG, AP, Apace, Vpace, Internal
FREQUENCY WEANING 1:1, 1:2, 1:3
HELIUM CYLINDER SIZE
Pressure2,216 psi
Volume, l 93, refillable tank
Battery type Rechargeable lead acid
BALLOON CAPACITY, CCRechargeable lead acid
EquivalentAbiomed, Datascope, Teleflex/Arrow
OPERATING DURATION, HR. 2
PUMP RATE LIMIT, BPM200
TIMING SETTINGSAuto-timing adjust for rate changes and arrhythmias based on beat-beat analysis, R-wave deflation when arrhythmia detected; manual mode to manually adjust inflation and deflation timing; inflation range: 20-80% (ECG, A-Pacer, V-Pacer, Internal); 0-35% (AP
INFLATION VOLUME ADJUST Increments, cc10% increment of balloon volume for augmentation changes
ALERTS
Balloon overinflate Yes
POWER FAILUREYes
Other Augment alarm selectable
Temporary alarm silencing Yes, realarms every minute, visual alarm continues
Trigger loss/change Yes
Balloon leak Yes
Device malfunction Yes
Low battery Yes
Vacuum/pressure loss Yes
Over pressure Yes
Loading failure Yes
Balloon disconnect Yes
CHARGING DURATION, HR<16 hr
CONFIGURATIONStand-alone, portable, combination IABP/VAD console
OTHER ATTRIBUTES Console supports 2 different pneumatically driven ABIOMED ventricular assist devices (BVS 5000 blood pump and AB5000 Ventricle for left, right or biventricular support) to provide up to 6 L/min; tethered and retractable 8 MHz handheld doppler probe; 360
FDA CLEARANCE Yes
CE MARK Yes
MARKETING REGION Worldwide
Mai puteti fi interesat de urmatoarele specificatii ale producatorului:
Disclaimer
Informatiile de pe aceasta pagina pot fi partiale sau incomplete si pot contine erori sau imprecizii.
MedWOW isi declina orice obligatie, responsabilitate sau garantie privitoare la sau in legatura cu orice informatie de pe aceasta pagina.

Pompa intra-aortica cu balon, ABIOMED, iPulse
View More Items